Skip to main content

Table 1 Analytical performance data of the calibration graphs for the determination of benzophenone by the proposed methods

From: Derivative spectrophotometric analysis of benzophenone (as an impurity) in phenytoin

Parameter

First derivative method (1D)

Third derivative method (3D)

Ratio derivative method (1DD)

Wavelength λ (nm)

257.6

263.2

272

Concentration range (μg/mL)

1-10

1-10

1-10

Intercept (a)

20.65 × 10-3

1.98 × 10-3

-0.67

Slope (b)

0.27

0.12

9.24

Correlation coefficient (r)

0.9999

0.9999

0.9999

Standard deviation of the residuals, Sy/x

4.77 × 10-3

5.14 × 10-3

0.21

Standard deviation of the intercept, Sa

3.71 × 10-3

3.99 × 10-3

0.16

Standard deviation of the slope, Sb

0.61 × 10-3

0.66 × 10-3

0.02

Relative standard deviation, % RSD

0.83

1.10

0.97

Percentage error, % Error

0.34

0.45

0.39

Limit of detection (LOD) (μg/mL)

0.04

0.11

0.06

Limit of quantitation (LOQ) (μg/mL)

0.13

0.34

0.18

D A1% (dl.g-1.cm-1)a

2655

1181

9.24 × 104

D ε (L.mol-1.cm-1)b

4.83 × 104

2.15 × 104

1.68 × 107

  1. a Specific absorbance of the derivative mode.
  2. b Molar absorptivity of the derivative mode.