Table 1 System suitability parameter for the optimum chromatographic condition for separation of CNB and its H-impurity

From: Development of a stability indicating high-performance liquid chromatography method for determination of cenobamate: study of basic degradation kinetics

Condition

Retention time

Number of theoretical plates

USP Tailing factor

Resolution

Capacity factor (k)

(CNB)

CNB

H-impurity

CNB

H-impurity

CNB

H-impurity

Optimum condition

Thermo, Hypersil BDS C18 column (150 × 4.6 mm, 5 μm), mobile phase: Methanol: purified water 50:50, %v/v

3.2

4.14

7591

8369

1.06

1.04

5.8

2.56