Skip to main content

Table 5 Assay validation sheet of FPV, CEF and MFX by the proposed models

From: Green chemometric-assisted UV-spectrophotometric methods for the determination of favipiravir, cefixime and moxifloxacin hydrochloride as an effective therapeutic combination for COVID-19; application in pharmaceutical form and spiked human plasma

Validation parameters

PLS

GAPLS

FPV

CEF

MFX

FPV

CEF

MFX

Wavelength (nm)

200–400

200–400

200–400

200–400

200–400

200–400

Linearity range (μg/mL)

1–15

2–15

1–10

1–15

1–15

0.5–10

Slopea

0.993

1.005

1.001

0.992

1.013

0.999

Intercepta

0.039

− 0.030

− 0.008

0.045

-0.069

0.007

Coefficient of determination (r2)a

0.9969

0.9945

0.9988

0.9975

0.9960

0.9996

LOD (μg/mL)b

0.291

0.383

0.177

0.261

0.323

0.104

LOQ (μg/mL)b

0.883

1.160

0.536

0.789

0.979

0.314

RMSECc

0.089

0.097

0.054

0.080

0.084

0.032

RMSEPd

0.137

0.173

0.118

0.133

0.131

0.097

RRMSEPe

2.741

3.458

2.356

2.205

2.904

2.205

BCMSEPf

0.014

0.029

0.013

0.012

0.021

0.012

RMSECVg

0.110

0.160

0.111

0.091

0.106

0.045

Accuracy (% R)h

99.49

100.10

100.12

100.59

98.95

99.49

Precision (% RSD)i

 Repeatability

1.483

1.281

1.452

1.197

1.260

1.184

 Intermediate precision

1.141

1.766

1.388

1.691

1.474

1.427

  1. aData of the straight line plotted between predicted concentrations versus actual concentrations of the calibration set
  2. bThe LOD and LOQ calculations are based on the net analyte signals
  3. cRoot-mean-square error of calibration
  4. dRoot-mean-square error of prediction
  5. eRelative root-mean-square error of prediction
  6. fBias-corrected mean square error of prediction
  7. gRoot mean square error of cross-validation
  8. hAverage of nine determinations (three concentrations repeated three times)
  9. i%RSD of nine determinations (three concentrations repeated three times)