Skip to main content

Table 7 Robustness of the developed method at the change of flow rate and mobile phase for the Bromoxynil and MCPA

From: Simultaneous determination of bromoxynil and MCPA in commercial samples and raw materials using reversed phase high performance liquid chromatography

(a) Robustness of the developed method at the change of flow rate and mobile phase for the Bromoxynil

Sample No

Change of Flow Rate

Change of Mobile Phase

Peak area at 1.3 mL/min

Peak area at 1.5 mL/min

Peak area at 1.7 mL/min

Methanol: Water

95: 05

Methanol: Water

90: 10

Methanol: Water

85: 15

01

16899261

14648612

13095952

11794631

14648612

19704492

02

16893790

14660667

13112501

11793598

14660667

19731947

03

16880153

14661086

13109821

11759541

14661086

19757886

04

16901669

14651265

13109991

11777437

14651265

19760224

05

16927612

14615510

13108926

11726960

14615510

19762280

Mean

16900497

14647428

13107438

11770433

14647428

19743366

Standard deviation

17299

18687

6557

28195

18687

24972

% RSD

0.10%

0.13%

0.05%

0.24%

0.13%

0.13%

(b) Robustness of the developed method at the change of flow rate and mobile phase for the MCPA

Sample No

Change of Flow Rate

Change of Mobile Phase

Peak area at 1.3 mL/min

Peak area at 1.5 mL/min

Peak area at 1.7 mL/min

Methanol: Water

95: 05

Methanol: Water

90: 10

Methanol: Water

85: 15

01

14290800

12455385

11060939

9951291

12455385

16818457

02

14279376

12380559

11058719

9950908

12380559

16843060

03

14267751

12381573

11057852

9921681

12381573

16875223

04

14294707

12469138

11051754

9910312

12469138

16883146

05

14342259

12459192

11047533

9899583

12459192

16904524

Mean

14294979

12429169

11055359

9926755

12429169

16864882

Standard deviation

28451

44200

5544

23557

44200

34072

% RSD

0.20%

0.36%

0.05%

0.24%

0.36%