Skip to main content

Table 4 Repeatability and Reproducibility of the developed method for the Bromoxynil and MCPA

From: Simultaneous determination of bromoxynil and MCPA in commercial samples and raw materials using reversed phase high performance liquid chromatography

(a) Repeatability of the developed method for the Bromoxynil and MCPA

Sr. #

Observations

Peak area (Bromoxynil)

Peak area (MCPA)

Solution of standard

Solution of sample

Solution of standard

Solution of sample

1

Reading 1

10886429

14648612

8370857

12455385

2

Reading 2

10892272

14660667

8375569

12380559

3

Reading 3

10902506

14661086

8394953

12381573

4

Reading 4

10891796

14651265

8375286

12469138

5

Reading 5

10889786

14615510

8373956

12459192

6

Average

10892558

14647428

8378124

12429169

7

Standard Deviation

6019

18687

9592

44200

8

RSD%

0.06%

0.13%

0.11%

0.36%

(b) Reproducibility of the developed method for the Bromoxynil and MCPA

Sr. #

Observations

HPLC—20AT

HPLC—10AT

Bromoxynil

MCPA

Bromoxynil

MCPA

1

Reading 1

14648612

12455385

9263118

7598232

2

Reading 2

14660667

12380559

9213510

7560822

3

Reading 3

14661086

12381573

9223018

7560822

4

Reading 4

14651265

12469138

9227850

7573807

5

Reading 5

14615510

12459192

9238788

7576635

6

Average

14647428

12429169

9233257

7574064

7

Standard Deviation

18687

44200

19015

15342

8

RSD %

0.13%

0.36%

0.21%

0.20%