Skip to main content

Table 3 Trueness of the developed method for the Bromoxynil and MCPA

From: Simultaneous determination of bromoxynil and MCPA in commercial samples and raw materials using reversed phase high performance liquid chromatography

(a) Trueness of the developed method for the Bromoxynil

Conc. (mg/L)

Mean peak area of standarda

Mean peak area of samplea

Observed yield (mg/L)

Percentage recovery (%)

150

6490553

6461836

149.3

99.6%

200

8481149

8460868

199.5

99.8%

250

10236719

10202255

249.2

99.7%

300

11811751

11779311

299.2

99.7%

350

13282454

13231410

348.7

99.6%

(b) Trueness of the developed method for the MCPA

Conc. (mg/L)

Mean peak area of standarda

Mean peak area of samplea

Observed yield (mg/L)

Percentage recovery (%)

150

4806238

4772312

149.0

99.3%

200

6328977

6288986

198.7

99.4%

250

7811343

7768557

248.6

99.5%

300

9187722

9125635

298.0

99.3%

350

10551859

10476843

347.5

99.3%

  1. aAverage of three replicate measurements