Skip to main content

Table 2 Regression parameters of the proposed smartphone-based colorimetric POCT method for BAC determination in urine

From: Smartphone based colorimetric point-of-care sensor for abused drugs: case of baclofen determination in urine

Parameter

BAC

Linearity range (mmol L−1 urine)

0.02–0.21

LLOQ (mmol L−1 urine)

0.02

Determination coefficient (r2)

0.9996

Slope ± Sb

5.66 ± 0.055

Intercept ± Sa

 − 0.047 ± 0.007

Sb%

0.977

S y/x

0.007

F value

10,476.36

Significance F

5.463 × 10–8