Skip to main content

Table 1 Optimization of parameters affecting smartphone based colorimetric analysis: a) Color scale channel and imaging background and b) distance between smartphone camera and sample cuvette

From: Smartphone based colorimetric point-of-care sensor for abused drugs: case of baclofen determination in urine

a

  

Sa

r2a

RGB scale channel

Red

0.292

0.7274

Green

0.874

0.9780

Blue

11.52

0.9989

Imaging background

White

11.52

0.9989

Black

4.389

0.9968

b

Distance (cm)

3.0

5.0

7.0

9.0

11.0

RSD % b

14.624

7.003

3.915

7.049

3.972

  1. aSensitivity (Slope) and r2 values were calculated using BAC concentrations of 0.02–0.21 mmol L−1