Skip to main content

Table 3 The robustness study for estimation of PCP with N-CQDs

From: Green microwave quantum dots as luminescent probes for quantifying prucalopride: consistency of content and application to pharmacokinetic studies

Variations

PCP (100 ng mL-1)

% Recoverya ± RSD

Optimum condition

101.55 ± 0.94

 1- Value of pH (BR buffer)

 6

99.71 ± 1.06

 6.4

99.60 ± 0.74

2- Volume of buffer (mL)

 0.75

99.79 ± 0.50

 1.25

99.30 ± 0.91

3- N@CQDs concentration (mg mL-1)

 0.14

99.75 ± 0.44

 0.16

99.94 ± 0.92

4- Reaction time (min)

 6

99.91 ± 0.59

 10

99.90 ± 0.31

  1. aMean of three determinations