Skip to main content

Table 2 The accuracy, intra-day precision and inter-day precision validations of the creative approach

From: Green microwave quantum dots as luminescent probes for quantifying prucalopride: consistency of content and application to pharmacokinetic studies

Sample number

Taken

(ng mL−1)

Found

(ng mL−1)

% Recovery* ± RSD

1

3.0

3.01

100.33 ± 0.60

2

50.0

50.11

100.22 ± 0.82

3

100.0

99.98

99.98 ± 0.34

4

150.0

152.28

101.52 ± 1.04

5

200.0

205.44

102.72 ± 1.41

Intra-day precision

50.0

50.23

100.46 ± 0.53

100.0

100.12

100.12 ± 0.66

150.0

151.20

100. 80 ± 0.61

Inter-day precision

50.0

49.49

98.98 ± 0.85

100.0

99.95

99.95 ± 1.38

150.0

150.13

100.08 ± 0.90

  1. *Average of three determinations. RSD Relative standard deviation