Skip to main content

Table 5 Precision results of the proposed HPLC–UV method

From: Development and validation of an eco-friendly HPLC–UV method for determination of atorvastatin and vitamin D3 in pure form and pharmaceutical formulation

Drug

Added concentration (µg/ml)

Intra-day precision

Inter-day precision

Mean*% ± SD

RSD%

Mean*% ± SD

RSD%

Atorvastatin calcium

10

98.44 ± 0.27

0.28

98.79 ± 0.47

0.48

20

101.26 ± 0.89

0.88

102.06 ± 1.04

1.02

40

99.49 ± 0.31

0.32

100.14 ± 0.74

0.74

Vitamin D3

2

98.41 ± 0.49

0.50

98.77 ± 0.59

0.60

4

101.43 ± 0.45

0.44

101.91 ± 0.62

0.60

8

100.21 ± 0.32

0.32

100.51 ± 0.39

0.39

  1. *Mean of three determinations