Skip to main content

Table 3 Accuracy results of the proposed HPLC–UV method

From: Development and validation of an eco-friendly HPLC–UV method for determination of atorvastatin and vitamin D3 in pure form and pharmaceutical formulation

Drug

Pure added (µg/ml)

Pure found (µg/ml)

Pure recovery* (R%)

Atorvastatin calcium

10

9.82

98.19

20

19.87

99.37

40

40.33

100.84

Vitamin D3

2

1.96

97.97

4

4.06

101.46

8

8.04

100.46

  1. *Mean of three determinations