Skip to main content

Table 4 Bioanalytical validation parameters for quantitative analysis of FPV and RDV by the proposed methods

From: Quantitative analysis of two COVID-19 antiviral agents, favipiravir and remdesivir, in spiked human plasma using spectrophotometric methods; greenness evaluation

Parameters

RD

1DD

FPV

RDV

FPV

RDV

LLOQ (µg/mL)

1.5

1.4

1.5

1.4

Linearity range (µg/mL)

1.5–24

1.4–24

1.5–24

1.4–24

Accuracy (%R)a

95.74

96.53

96.89

95.02

Repeatability precision (RSD)b

1.567

1.897

1.418

1.658

Intermediate precision (RSD)b

1.774

2.001

1.704

1.911

Matrix effect (Mean ± RSD)

96.30 ± 1.720

95.14 ± 1.936

96.25 ± 1.340

95.81 ± 1.437

  1. a The mean percentage recovery of tripicate determination of four quality control samples
  2. b RSD of tripicate determination of four quality control samples