Table 3 Application of the proposed method for the determination of OMG, MET, and EZT in pure forms

From: Analytical quality-by-design approach for development and validation of HPLC method for the simultaneous estimation of omarigliptin, metformin, and ezetimibe: application to human plasma and dosage forms

Parameter

OMG

MET

EZT

Conc. taken

(µg/mL)

%Founda

Conc. taken

(µg/mL)

%Founda

Conc. taken

(µg/mL)

%Founda

 

0.2

97.5

0.5

97.2

0.1

100.7

0.25

100.44

1.0

99.54

0.2

104.15

0.3

96.8

2.5

98.2

0.3

100.2

0.5

102.82

5.0

101.21

0.4

99.2

0.75

100.97

10.0

99.83

0.5

100.52

1.0

98.83

15.0

100.97

1.0

97.67

1.5

100.79

20.0

99.06

1.5

100.93

2.0

99.65

25.0

100.25

2.0

100.01

Mean

 

99.73

 

99.53

 

100.42

± SD

1.97

1.36

1.83

%RSD

1.972

1.365

1.825

% Error

0.695

0.48

0.648

  1. a Mean of three determinations.