Table 1 Summary of quantitative parameters and statistical data using the proposed methods

From: A novel spectrophotometric approach relies on a charge transfer complex between atomoxetine with quinone-based π-acceptor. Application to content uniformity test

Parameters

Chloranil method

DDQ method

Linear range (µg mL− 1)

30–320

10–80

Intercept (a)

0.0259

− 0.0134

Standard deviation of intercept (SD a)

0.004007

0.00637

Slop (b)

0.0027

0.0134

Standard deviation of slope (SD b)

0.000021

0.00013

Correlation coefficient (r)

0.9998

0.9997

Determination coefficient (r 2 )

0.9996

0.9995

Molar absorptivity Ɛ (L / mol cm )

858.64

3805.107

Limit of quantitation (LOQ), µg mL − 1

14.84

4.75

Limit of detection (LOD), µg mL − 1

4.89

1.56

  1. Number of determinations were 8 for both methods