Skip to main content

Table 6 Precision results of 23 elements in the standard serum from seronorm™ trace element serum L1 and L2 (n = 3)

From: A simple, rapid method for simultaneous determination of multiple elements in serum by using an ICP-MS equipped with collision cell

Element

L1 (RSD%)

L2 (RSD%)

 

Intra-day

Inter-day

Intra-day

Inter-day

Mg

1.92

4.63

1.21

1.25

Al

4.80

6.89

2.53

4.27

V

3.70

6.86

0.91

6.02

Cr

2.06

3.38

1.73

2.28

Mn

5.39

7.62

1.55

3.42

Fe

2.44

8.17

0.63

2.60

Co

2.62

5.70

1.07

2.82

Ni

1.70

8.94

1.23

1.42

Cu

2.27

3.28

1.81

2.67

Zn

1.92

4.65

1.89

2.98

As

1.17

7.41

5.24

11.97

Se

1.98

9.55

1.47

5.73

Rb

2.34

2.58

1.75

2.35

Sr

1.55

4.49

1.56

2.45

Mo

1.16

6.80

0.65

4.56

Cd

6.13

10.65

6.79

7.11

Sn

4.18

12.19

4.32

5.78

Sb

1.52

4.05

1.42

2.00

Ba

1.71

2.40

2.58

2.68

W

2.81

11.30

2.39

3.58

Tl

1.74

3.96

1.36

1.41

Pb

9.12

11.05

4.07

10.52

U

1.83

1.92

2.19

2.52

  1. RSD relative standard deviation