Skip to main content

Table 1 Electrochemical response characteristics of the fabricated sensors

From: Determination of Clomipramine using eco-friendly solid-contact ionophore-doped potentiometric sensor

Parameter

C-SPE/ISM(CX4)

C-SPE/Gr-NC/ISM(CX4)

Linear range

1 × 10− 2 to 1 × 10− 5.3 M

1 × 10− 2 to 1 × 10− 5.3 M

Slope (mV/decade)

-54.06

-55.42

Response time (s)

15

10

Working pH range

2–6

2–6

Stability (weeks)

4

4

LOD

1 × 10− 5.39 M

1 × 10− 5.52 M