Skip to main content

Table 7 System suitability testing parameters of the developed methods

From: Determination of flutamide and two major metabolites using HPLC–DAD and HPTLC methods

Parameters

HPTLC method

HPLC method

Flu-1

Flutamide

Flu-3

Flu-3

Flu-1

Flutamide

Rf (for HPTLC) or Rt (for HPLC)

0.48 ± 0.01

0.6 ± 0.01

0.74 ± 0.02

4.01 ± 0.06

6.00 ± 0.01

8.96 ± 0.02

Peak symmetry

1.00

1.00

0.94

1.30

1.25

1.10

Selectivity (α)

 Plasma

9.14

1.23

1.24

3.54

1.71

1.62

 Urine

9.8

  

1.71

  

Resolution (Rs)

 Plasma

10.18

2.00

2.24

4.60

5.24

5.60

 Urine

9.58

  

2.77

  

Capacity factor (α)

1.04

0.64

0.32

2.34

4.00

6.48

Number of theoretical plates (N)

 

3059.22

2515.28

3698.29

Height equivalent to theoretical plate (H) (in cm)

 

0.0049

0.0060

0.0041