Skip to main content

Table 2 Precision and accuracy for quality control standard of stanazolol

From: A validated stability-indicating TLC-densitometric method for the determination of stanozolol in pharmaceutical formulations

 

Analyst 1

Analyst 2

Intra-day

Inter-day

 

Conc.

Found*

R.S.D.

Accuracy

Found*

R.S.D.

Accuracy

Found*

R.S.D.

Accuracy

(ng)

(ng)

%

(%)

(ng)

%

(%)

(ng)

%

(%)

300

303.56 ± 5.00

1.65

98.81

306.63± 5.52

1.80

97.79

300.62± 1.452

0.48

99.79

500

503.33 ± 1.06

0.21

99.33

504.49± 3.38

0.67

99.10

502.63± 0.78

0.15

99.47

700

705.80 ± 1.50

0.21

99.17

703.85± 1.14

0.16

99.45

703.04± 0.71

0.10

99.56

  1. *Values are mean of n = 3.