Skip to main content

Table 1 Effect of experimental parameters on the number of theoretical plates of the studied compounds

From: Simultaneous determination of sulpiride and mebeverine by HPLC method using fluorescence detection: application to real human plasma

Parameter

No. of theoretical plates (N)

Resolution (Rs)

  

MEB

SUL

MEB

SUL

Ratio of organic modifier A/B

40/60

1199

2521

3.20

0.345

 

45/55

3407

4691

6.00

3.95

 

50/50

2867

1824

5.75

1.77

 

55/45

1658

911

3.97

1.20

 

60/40

1255

1348

3.97

1.46

 

70/30

1946

834

4.90

6.19

 

75/25

1487

825

4.20

0.99

pH

3.0

2687

1778

5.04

1.55

 

3.5

1138

1510

3.50

1.52

 

4.0

3462

4659

6.16

3.70

 

4.5

1716

2366

5.10

3.10

 

5.0

1345

1630

5.33

3.10

Ionic strength of phosphate buffer [M]

0.0075

2935

2315

5.08

2.24

 

0.01

3451

4632

6.16

3.72

 

0.02

1614

1251

5.60

2.62

 

0.03

1982

1150

6.20

2.50

 

0.04

1394

950

5.30

2.40

Flow rate mL/min

0.50

1297

1783

3.6

1.5

 

0.75

2356

3252

5.20

2.50

 

1.00

3442

4621

6.20

3.90

 

1.20

1864

2534

4.75

3.10

 

1.50

1534

1963

5.80

3.40