Table 2 Precision and accuracy in the assay determination of FEX using the proposed LC method

From: Development of validated stability-indicating chromatographic method for the determination of fexofenadine hydrochloride and its related impurities in pharmaceutical tablets

Day of analysis

Spiked concentration

(μg/ml)

Mean recovery (%) ± RSDa

Repeatability (Intra-day precision)

Day 1

0.1

101.5 ± 0.89

 

10

99.4 ± 0.49

 

50

101.2 ± 0.67

Day 2

0.1

103.8 ± 0.99

 

10

100.2 ± 0.67

 

50

101.8 ± 0.89

Day 3

0.1

102.4 ± 0.1.2

 

10

99.4 ± 0.59

 

50

100.7 ± 0.55

Intermediate precision (Inter-day precision)

 

0.1

102.5 ± 1.13

 

10

99.6 ± 0.46

 

50

101.23 ± 0.54

  1. a Mean recovery (%) ± RSD of three determinations.