Skip to main content

Table 5 Detection results of the standard serum for quality control of seronorm™ trace element serum L2

From: A simple, rapid method for simultaneous determination of multiple elements in serum by using an ICP-MS equipped with collision cell

Element

Analytical value of the standard serum (μg/L)

Analytical uncertainty of the standard serum (95% confidence interval) (μg/L)

Measured value (mean ± SD) (μg/L)

Mga

33.900

27.1–40.7

39.958 ± 1.966

Al

120.000

96–144

142.884 ± 9.630

V

1.100

1.211 ± 0.072

Cr

5.700

4.0–7.5

5.164 ± 0.418

Mn

14.500

11.6–17.4

16.243 ± 1.563

Fea

2.15

1.72–2.58

2.202 ± 0.151

Co

3.050

2.13–3.97

2.982 ± 0.274

Ni

9.000

7.9–11.9

10.797 ± 0.808

Cua

1.925

1.538–2.312

1.931 ± 0.037

Zna

1.532

1.223–1.840

1.792 ± 0.039

As

0.380

0.357 ± 0.036

Se

136.000

95–176

107.691 ± 7.299

Rb

8.700

9.109 ± 0.419

Sr

110.000

113.189 ± 3.424

Mo

1.210

1.176 ± 0.043

Cd

0.140

0.160 ± 0.012

Sn

0.250

0.356 ± 0.024

Sb

15.000

18.124 ± 0.339

Ba

139.135

137.563 ± 4.108

W

0.065

0.062 ± 0.006

Tl

0.108

0.126 ± 0.002

Pb

0.660

0.732 ± 0.065

U

0.359

0.377 ± 0.014

  1. amg/L