Skip to main content

Table 4 Detection results of the standard serum for quality control of seronorm™ trace element serum L1

From: A simple, rapid method for simultaneous determination of multiple elements in serum by using an ICP-MS equipped with collision cell

Element

Analytical value of the standard serum (μg/L)

Analytical uncertainty of the standard serum (95% confidence interval) (μg/L)

Measured value (mean ± SD) (μg/L)

Mga

16.8

13.4–20.1

19.363 ± 0.878

Al

50.400

25.2–75.7

56.284 ± 3.598

V

1.100

1.266 ± 0.089

Cr

2.180

1.30–3.05

2.091 ± 0.197

Mn

9.900

7.9–11.9

11.202 ± 0.860

Fea

1.47

1.17–1.77

1.545 ± 0.1288

Co

1.120

0.67–1.57

1.094 ± 0.069

Ni

5.640

3.38–7.90

6.452 ± 0.569

Cua

1.066

0.852–1.281

1.123 ± 0.028

Zna

1.057

0.844–1.269

1.229 ± 0.043

As

0.400

0.315 ± 0.019

Se

86.000

51–120

67.786 ± 6.477

Rb

4.400

4.679 ± 0.125

Sr

95,99

99.650 ± 4.188

Mo

0.760

0.804 ± 0.039

Cd

0.130

0.165 ± 0.015

Sn

0.250

0.366 ± 0.043

Sb

10.400

12.941 ± 0.374

Ba

189.000

183.292 ± 3.637

W

0.058

0.074 ± 0.008

Tl

0.090

0.110 ± 0.004

Pb

0.400

0.539 ± 0.057

U

0.302

0.337 ± 0.012

  1. amg/L