Skip to main content

Table 1 iCAP Qc ICP-MS daily working parameters

From: A simple, rapid method for simultaneous determination of multiple elements in serum by using an ICP-MS equipped with collision cell

Instrumental parameter

Values

Plasma gas

15 L/min

Auxilliary gas

0.8 L/min

nebulizer gas flow

1.02 L/min (optimized daily)

He collins gas

4.60 ml/min (optimized daily)

RF power

1550 W

Detector mode

Dual

Channel

3

Spacing

0.1

Resolution

Normal

Measurement mode

KED

Standard curve type

Simple linear

Number of sweeps

20

Main runs

3

Dwell time

0.01 s

  1. KED kinetic energy discrimination