Skip to main content

Table 5 Statistical comparison of the results of the proposed sensors with the reported method

From: Fabrication of novel electropolymerized conductive polymer of hydrophobic perfluorinated aniline as transducer layer on glassy carbon electrode: application to midazolam as a model drug of benzodiazepines

Parameters

PTFANI-free sensor

PTFANI-sensor

Reported methoda

n

9

9

9

Mean

99.13

99.89

100.69

SD

1.47

1.13

0.92

Variance

2.16

1.28

0.85

t-test (2.12)b

0.968

1.64

 

F (3.44)b

2.541

1.51

 
  1. aReported HPLC method: C18 column, using a mobile phase: CAN-phosphate buffer (pH 3.3) (30:70 v/v) at a flow rate of 1.0 mL/min and UV detection at 220 nm