Table 3 Validation parameters of the developed electrochemical methods

From: Determination of Clomipramine using eco-friendly solid-contact ionophore-doped potentiometric sensor

Parameter

C-SPE/ISM(CX4)

C-SPE/Gr-NC/ISM(CX4)

Linearity

Slopea

-54.06

-55.42

Intercept

411.23

382.11

R2

0.9932

0.9936

Range

1 × 10− 2 − 1 × 10− 5.3 M

Accuracyb

100.60 ± 1.75

101.05 ± 1.10

Precision

Repeatabilityc

Intermediate precisionc

0.668

1.862

0.822

1.516

  1. a Average of three determinations
  2. b Accuracy results were done for (5 × 10− 5, 5 × 10− 4 and 5 × 10− 3.5 M) of CLM.
  3. c RSD for repeatability and intermediate precision were done for (10− 5, 10− 4 and 10− 3 M) of CLM.