Parameters | MOL |
---|
Concentration range | 100–2000 (ng/mL) |
Correlation coefficient | 0.9999 |
Slope | 0.0006 |
Intercept | 0.0456 |
S.D of intercepta | 0.0056 |
LODb | 30 (ng/mL) |
LOQb | 90.77 (ng/mL) |
Accuracy |
Mean ± SD | 100.89 ± 0.60 |
RSD% | 0.60 |
Er%c | 0.89 |
Intra-day precisiond |
Mean ± SD | 101.16 ± 1.12 |
RSD% | 1.11 |
Er%c | 1.16 |
Inter-day precisione |
Mean ± SD | 100.58 ± 0.95 |
RSD% | 0.94 |
Er%c | 0.58 |
- aStandard deviation of intercept
- bLOD = (SD of the response/slope) × 3.3; LOQ = (SD of the response/slope) × 10
- cRelative error percentage
- dThe intra-day analysis, average of three different concentrations of MOL (500, 1000, and 1500 ng/mL) repeated three times within the day
- eThe inter-day analysis, average of three different concentrations of MOL (500, 1000, and 1500 ng/mL) repeated three times in three consecutive days