Skip to main content

Table 9 Assay results for the determination of the ASP, ATR and RAM drugs in synthetic mixtures in ratios of 10:4:1 (w/w) by the proposed HPLC method

From: Micellar eco-friendly HPLC method for simultaneous analysis of ternary combination of aspirin, atorvastatin and ramipril: application to content uniformity testing

Compound

Proposed method

Comparison method [9]

Amount taken, (µg/mL)

% Found, %

% Found, %

ASP

100

99.14

101.02

140

101.10

100.45

200

99.71

100.58

Mean ± SD

 

99.98 ± 1.01

100.35 ± 0.73

t-test

1.15 (2.78)

 

F- test

11.39 (19.00)

ATR

40

99.33

99.67

56

100.84

100.28

80

99.97

99.78

Mean ± SD

 

100.05 ± 0.76

99.91 ± 0.33

t-test

 

0.29 (2.78)

 

F-test

 

5.43 (19.00)

RAM

10

101.57

100.03

14

98.59

98.56

20

100.31

101.46

Mean ± SD

 

100.16 ± 1.50

100.017 ± 1.45

t-test

 

0.12 (2.78)

 

F-test

 

1.06 (19.00)