Skip to main content

Table 1 The custom experimental design architecture for the levels and components of the studied sensors and the observed responses for each sensor

From: Experimentally designed electrochemical sensor for therapeutic drug monitoring of Ondansetron co-administered with chemotherapeutic drugs

Sensor no

Factor 1

Factor 2

Factor 3

Response 1

Response 2

Response 3

Response 4

Plasticizer

Ion exchanger

Ionophore

Slope

r

pLOQ

\(\mathrm{log}{K}_{OND, {Na}^{+}}^{pot}\)

1

−1

NPOE

2

RK

−2

BCD

53.55

0.9984

4.57

−2.50

2

1

DOP

2

RK

−1

HPBCD

53.10

0.9992

4.57

−3.71

3

−1

NPOE

−1

PT

−1

HPBCD

62.79

0.9983

4.36

−4.22

4

−1

NPOE

1

TKS

1

CX4

59.70

0.9972

5.32

−2.96

5

1

DOP

−1

PT

0

CMBCD

62.56

0.9986

4.79

−3.44

6

1

DOP

1

TKS

2

CX8

56.49

0.9976

5.03

−3.56

7

−1

NPOE

0

PM

−1

HPBCD

59.05

0.9980

4.01

−4.23

8

−1

NPOE

−2

TPB

2

CX8

59.05

0.9965

4.57

−2.11

9

−1

NPOE

−1

PT

−2

BCD

66.41

0.9988

5.04

−2.88

10

1

DOP

−1

TPB

−2

BCD

60.78

0.9960

5.04

−2.86

11

−1

NPOE

2

RK

2

CX8

59.05

0.9981

4.57

−2.96

12

−1

NPOE

0

PM

0

CMBCD

56.04

0.9994

4.52

−3.12

13

−1

NPOE

1

TKS

0

CMBCD

60.15

0.9980

5.04

−2.94

14

−1

NPOE

−2

TPB

1

CX4

66.75

0.9971

5.32

−2.25

15

1

DOP

0

PM

1

CX4

50.25

0.9986

4.97

−3.26