Table 2 Accuracy and precision for the analysis of CAB in human plasma samples

From: A validated LC–MS/MS method for analysis of Cabergoline in human plasma with its implementation in a bioequivalent study: investigation of method greenness

 

LLOQ

2 pg/mL

Low QC

6 pg/mL

Med QC (A)

20 pg/mL

Med QC (B)

60 pg/mL

High QC

160 pg/mL

Day 1 intra-day accuracy (N = 6)

Equation (1/X)

Y = 1.299784 X − 0.017059 (R2 = 0.9978)

STD

0.098

0.389

0.423

2.350

2.535

CV%

4.67

6.77

2.02

3.75

1.59

Accuracy%

105.38%

95.88%

104.64%

104.42%

99.71%

Day 2 intra-day accuracy (N = 6)

Equation (1/X)

Y = 1.078651X − 0.030379 (R2 = 0.9968)

STD

0.225

0.123

0.263

2.203

7.607

CV%

11.917

2.118

1.400

3.737

4.941

Accuracy%

94.31

97.15

93.75

98.24

96.22

Day 3 inter-day results (N = 18)

Equation (1/X)

Y = 1.238189*X + 0.027894 (R2 = 0.9944)

STD

0.219

0.472

1.176

3.082

5.248

CV%

11.107

8.054

5.792

5.119

3.326

Accuracy%

98.46%

97.63%

101.54%

100.35%

98.60%

  1. *Results were excluded if they were out of the acceptance range (85–115%)