Skip to main content

Table 2 Validation: precision and accuracy data for thymoquinone in different biomatrixes

From: A bioanalytical UHPLC based method used for the quantification of Thymoquinone-loaded-PLGA-nanoparticles in the treatment of epilepsy

Biomatrix

Quality controls samples

Theoretical concentration

(ng mL−1) or

(ng g−1)

Intra-batch precision

Inter-batch precision

Recoveryc

(%)

Observed concentration

(ng mL−1) or (ng g−1) ± S.D.

Accuracya

(%)

Precisionb

(%C.V.)

Observed concentration

(ng mL−1) or (ng g−1) ± S.D.

Accuracya

(%)

Precisionb

(%C.V.)

Brain Homogenate

LOQQC

101.00

97.79 ± 1.38

96.82

1.41

96.63 ± 1.27

95.67

1.31

85.34

LQC

290.00

276.94 ± 3.64

95.50

1.31

274.56 ± 3.11

94.68

1.13

86.47

MQC

1000.00

979.64 ± 11.46

97.96

1.17

974.43 ± 11.17

97.44

1.15

87.61

HQC

2000.00

1985.22 ± 27.11

99.26

1.37

1982.42 ± 26.64

99.12

1.34

88.23

Lungs Homogenate

LOQQC

101.00

97.68 ± 0.934

96.71

0.96

98.28 ± 1.14

97.31

1.16

83.14

LQC

290.00

282.46 ± 3.75

97.40

1.33

279.99 ± 3.64

96.55

1.30

88.18

MQC

1000.00

976.49 ± 15.47

97.65

1.58

974.94 ± 14.99

97.49

1.54

87.23

HQC

2000.00

1981.44 ± 25.14

99.07

1.27

1980.47 ± 24.34

99.02

1.23

87.43

Plasma

LOQQC

101.00

99.64 ± 1.16

98.65

1.16

99.31 ± 1.01

98.33

1.02

88.32

LQC

290.00

281.45 ± 3.93

97.05

1.40

279.64 ± 3.81

96.43

1.36

86.93

MQC

1000.00

976.42 ± 18.69

97.64

1.91

972.25 ± 18.14

97.23

1.86

87.39

HQC

2000.00

1976.48 ± 26.18

98.82

1.32

1974.14 ± 25.22

98.71

1.28

87.24