Skip to main content

Advertisement

Table 9 Operational range and corrected operational range

From: Evaluation of an online, real-time, soft-photon ionisation time-of-flight mass spectrometer for mainstream tobacco smoke analysis

AnalyteRange, μg/puffCorrected range, μg/puff
Acetaldehyde8.1–81.611–110.9
Acetone3.6–35.64.2–41.4
1,3-Butadiene0.8–8.60.7–7.6
2-Butanone2–9.52.2–10.3
Benzene1.1–8.51.2–8.9
Isoprene10.9–66.79.3–56.8
Toluene2–13.22.0–13.3