Table 9 Operational range and corrected operational range

From: Evaluation of an online, real-time, soft-photon ionisation time-of-flight mass spectrometer for mainstream tobacco smoke analysis

Analyte

Range, μg/puff

Corrected range, μg/puff

Acetaldehyde

8.1–81.6

11–110.9

Acetone

3.6–35.6

4.2–41.4

1,3-Butadiene

0.8–8.6

0.7–7.6

2-Butanone

2–9.5

2.2–10.3

Benzene

1.1–8.5

1.2–8.9

Isoprene

10.9–66.7

9.3–56.8

Toluene

2–13.2

2.0–13.3