Skip to main content

Table 3 Analytical parameters of methylphenidate-PVC sensors

From: New potentiometric sensors for methylphenidate detection based on host–guest interaction

Parameter

Sensor 1

Sensor 2

Sensor 3

Slope, (mV/decade)

59.4 ± 0.5

51.37 ± 0.5

56.5 ± 0.5

Intercept, mV

220.45 ± 0.5

216.58 ± 0.5

248.17 ± 0.5

Calibration range

8 × 10−6–1 × 10−3

8.0 × 10−6–1 × 10−3

8.0 × 10−6–1 × 10−3

STE YX

0.816

2.85

0.816

SE slope

0.5574

2.02

0.5574

SE Intercept

2.357

8.25

2.357

Correlation coefficient, (r2)

0.999

0.999

0.996

Lower limit of quantification (LOQ), M

8 × 10−6

8 × 10−6

8 × 10−6

Lower limit of detection (LOD), M

7.5 × 10−6

7 × 10−6

7 × 10−6

Response time for 1 × 10−3 M solution, (S)

25 ± 0.5

25 ± 0.5

25 ± 0.5

Working pH range

4–8

4–8

4–8

  1. Sensor 1 (β-CD), Sensor 2 (γ-CD) and Sensor 3 (calaxirene)
  2. SE standard error