Skip to main content

Table 1 Analytical figures of merit for Zr-MOF-NH2 for extraction of CBZ

From: Determination of carbamazepine in urine and water samples using amino-functionalized metal–organic framework as sorbent

Parameter

Analytical feature

Linear Dynamic range (μg L−1)

0.1–50

R2 (determination coefficient)

0.9988

Repeatability (RSD%) (50 μg L−1)

2.5

Limit of detection (µg L−1)

0.04

Total extraction time (min)

≤ 12