Skip to main content

Table 6 Instrumental conditions for the determination of each element (FAAS technique)

From: Vertical distribution and analysis of micro-, macroelements and heavy metals in the system soil-grapevine-wine in vineyard from North-West Romania

Element

Wavelength

Slit

Correlation

Flame

Background

LOD*

LOQ**

(nm)

(nm)

coefficient

(2300°C)

correction

(mg/L)

(mg/L)

Ca

422.7

0.7

1.000000

Air-acetylene

-

0.092

0.306

Mg

285.2

0.7

1.000000

Air-acetylene

Deuterium

0.190

0.633

Fe

248.3

0.2

0.999972

Air-acetylene

Deuterium

0.110

0.366

Zn

213.9

0.7

0.999999

Air-acetylene

Deuterium

0.018

0.059

Cu

324.8

0.7

0.999979

Air-acetylene

Deuterium

0.017

0.056

Ni

232.0

0.2

0.999920

Air-acetylene

Deuterium

0.020

0.066

Co

240.7

0.2

0.999900

Air-acetylene

Halogen

0.120

0.399

Pb

283.3

0.2

0.999853

Air-acetylene

Deuterium

0.051

0.166

Cd

228.8

0.7

1.000000

Air-acetylene

Deuterium

0.028

0.093

Na

589.0

0.2

1.000000

Air-acetylene

-

0.012

0.039

  1. *Detection limit.
  2. **Quantification limit.
  3. - Not used background correction.