Device | Command | Parameter | Action |
---|---|---|---|
‘Loop Start (♯) 1 | |||
Next sample | Counter clockwise | ||
Peristaltic pump | Delay (s) | % 50 | |
Peristaltic pump off | 25 | ||
Detector | ON | ||
Syringe pump | Valve position IN | ||
Syringe pump | Set flow rate (μL s −1 ) | 100 | |
Syringe pump | Aspirate (μL) | 1500 | Pump filled with carrier |
Syringe pump | Delay until done | ||
Multiposition valve | Set valve position | 3 | |
Syringe pump | Set flow rate (μL s −1 ) | 100 | |
Syringe pump | Aspirate (μL) | 50 | (Luminol 5.0×10 −4 mol L −1 ) |
Multiposition valve | Set valve position | 2 | |
Syringe pump | Set flow rate (μL s −1 ) | 100 | |
Syringe pump | Aspirate (μL) | 50 | (Gold nanoparticles 1.0×10 −5 mol L −1 ) |
Multiposition valve | Set valve position | 5 | |
Syringe pump | Set flow rate (μL s −1 ) | 100 | |
Syringe pump | Aspirate (μL) | 30 | Sample (NLX) |
Syringe pump | Delay until done | ||
Multiposition valve | Set valve position | 4 | |
Syringe pump | Set flow rate (μL s −1 ) | 100 | |
Syringe pump | Aspirate (μL) | 30 | (Potassium ferricyanide1.0×10 −2 mol L −1 ) |
Syringe pump | Delay until done | ||
Multiposition valve | Detector | 7 | |
Syringe pump | Set flow rate (μL s −1 ) | 100 | |
PMT | Start scan | ||
Syringe pump | Empty | ||
Syringe pump | Delay until done | ||
PMT | Stop scans | ||
Refresh plat | |||
Loop end |