Skip to main content

Table 2 Response characteristics of moxifloxacin-PVC matrix membrane sensors

From: Potentiometric determination of moxifloxacin in some pharmaceutical formulation using PVC membrane sensors

Parameter

MOX-TPB

MOX-PM

MOX-PT

Slope, (mV/ decade)

53.0 × 0.5

54.5 × 0.5

55.0 × 0.5

Intercept, mV

137.8 × 0.5

143.03 × 0.5

141.78 × 0.5

Correlation coefficient, (r)

0.996

0.995

0.995

Lower limit of detection (LOD), M

4.0 × 10-6

5 × 10-6

5 × 10-6

Lower limit of quantification (LOQ), M

3.0 × 10-6

4.0× 10-6

4.0 × 10-6

Response time for 1 × 10 -3 M solution, s

30 × 0.5

30 × 0.5

30 × 0.5

Working pH range

6.0 - 9.0

6.0 - 9.0

6.0 - 9.0