Table 3 Descriptive statistics for the Bland and Altman test (95% confidence level, m = 3 independent samples)

From: Validation of an analytical method based on the high-resolution continuum source flame atomic absorption spectrometry for the fast-sequential determination of several hazardous/priority hazardous metals in soil

Element

Within-method standard deviation (mg/kg)

Between-method standard deviation (sB) (mg/kg)

Confidence interval of bias Δ ¯ ± t s B n (mg/kg)

Confidence intervals for limits of agreement Δ ¯ ± 1.96 s B ± t · s LL , UL (mg/kg)

 

ICP-OES

HR-CS-AAS

   

Ag (n = 25)

0.4

0.3

1.8

-0.5 ± 0.8

-4.1 ± 1.3

3.1 ± 1.3

Cd (n = 25)

1.0

0.8

2.7

-0.6 ± 1.1

-5.9 ± 1.8

4.8 ± 1.8

Co (n = 25)

0.7

0.7

1.4

0.2 ± 0.6

-2.6 ± 0.8

3.0 ± 0.8

Cr (n = 25)

1.3

1.4

3.9

0.9 ± 1.6

-6.9 ± 2.5

8.6 ± 2.5

Cu (n = 34) (<1000)a

30.1

18.7

33.4

4.8 ± 11.7

-60.8 ± 14.0

70.3 ± 14.0

Cu (n = 16) (1000–22000)a

260

246

390

-57 ± 208

-821 ± 258

707 ± 258

Ni (n = 25)

0.9

1.9

2.5

-0.4 ± 1.0

-5.3 ± 1.3

4.5 ± 1.3

Pb (n = 28) (<1000)a

29.6

22.9

42.5

4.9 ± 16.5

78.3 ± 20.8

88.2 ± 20.8

Pb (n = 14) (1000–10000)a

174

56

214

-36 ± 123

-455 ± 162

383 ± 162

Pb (n = 13) (10000–60000)a

712

494

1477

-243 ± 893

-3138 ± 1343

2652 ± 1343

Zn (n = 30) (< 1000)a

7.5

5.2

13.2

-3.3 ± 4.9

-29.1 ± 6.9

22.6 ± 6.9

Zn (n = 20) (> 1000)a

122

90

169

13 ± 79

-319 ± 100

     

345 ± 100