Skip to main content

Table 1 Limits of detection (3σ criterion) and quantification (mg/kg) in soil

From: Validation of an analytical method based on the high-resolution continuum source flame atomic absorption spectrometry for the fast-sequential determination of several hazardous/priority hazardous metals in soil

 

HR-CS-FAAS

ICP-OES

Limit of applicability of LR-LS-FAASc

 

λ (nm)

LODa

LOQb

λ (nm)

LODa

LOQb

 

Ag

328.068

0.18

0.54

328.068

0.12

0.36

-

Cd

228.802

0.14

0.42

214.438

0.05

0.15

>2

Co

240.206

0.36

1.08

228.615

0.15

0.45

>12

Cr

357.869

0.25

0.75

267.716

1.4

4.2

>12

Cu

324.754

0.09

0.27

324.754

0.15

0.45

>5

Ni

232.003

1.0

3.0

305.082

2.5

7.5

>12

Pb

217.000

1.4

4.2

220.351

2.5

7.5

>15

Zn

213.856

0.18

0.54

213.856

0.04

0.12

>2

  1. a – LOD calculated for an amount of 1.000 g soil mineralized and diluted to 100 ml solution.
  2. b – LOQ considered three times LOD.
  3. c – according to [53].