Table 4 Validation metrics results of the SIC-FL method

From: Native fluorescent detection with sequential injection chromatography for doping control analysis

Analytical character

Result

Resolution

8.2

Consumed mobile phase volume (mL)

4.5

Total analysis time (min)

4.7

Sample frequency (samples/h)

13

 

AML

FSM

Retention time (min)

0.85

1.62

Peak symmetry

1.00

0.89

Theoretical plates

671.55

937.95

Regression equation

PAa = 1.6387Cb + 566.28

PAa = 26.667Cb + 23.333

Correlation coefficient

0.9971

0.9987

Linear range (μg/mL)

0.1-4.0

2.0-10.0

Within-run precision (RSD, %)

1.98

2.02

Between-run precision (RSD, %)

0.57

2.81

Recovery in urine samples (%)

89

91

Recovery in tablets (%)

99.1 for 1.0 μg/mL

98.7 for 8.0 μg/mL

 

97.5 for 0.5 μg/mL

98.1 for 4.0 μg/mL

 

97.0 for 0.5 μg/mL

96.5 for 0.5 μg/mL

LOD (μg/mL)

0.012

0.470

LOQ (μg/mL)

0.060

1.500

  1. apeak area; bconcentration (μg/mL)