Skip to main content

Table 3 Regression data, LODs and LOQs for the three analytes tested in HPLC-UV chromatograms

From: Quality assessment on Polygoni Multiflori Caulis using HPLC/UV/MS combined with principle component analysis

Analyte

Calibration curvea

R 2

Linear range (mg/L)

LODb(ng/mL)

LOQc(ng/mL)

1

y = 29885 x - 210960

0.9990

0.404-252.5

98.08

404.03

2

y = 38332 x + 23549

0.9999

0.0443-138.325

44.26

180.56

3

y = 22565 x - 1352.2

0.9997

0.123-15.4

56.77

132.20

  1. 1: THSG.
  2. 2: emodin.
  3. 3: physcion.
  4. ay is the peak area in UV chromatograms, x is the concentration (mg/L) of the analyte.
  5. b LOD refers to the limit of detection, s/n = 3.
  6. c LOQ refers to the limit of quantification, s/n = 10.