Table 6 Application of the proposed method for determination of the studied drugs in spiked and real human plasma

From: Simultaneous determination of sulpiride and mebeverine by HPLC method using fluorescence detection: application to real human plasma

Sample

Concentration taken (ng/mL)

Concentration found (ng/mL)

Recovery %

 

Veratic acid

SUL

Veratic acid

SUL

Veratic acid

SUL

1-Spiked plasma sample

      

a- Intra-day precision

10

10

10.32

9.62

103.19

96.23

 

25

50

24.34

51.04

97.36

102.09

 

50

100

50.28

99.50

100.55

99.50

X'

    

100.36

99.27

± S.D.

    

2.92

2.95

%RSD

    

2.91

2.95

%Error

    

1.68

1.69

b- Inter- day precision

50

100

50.56

95.32

101.12

95.30

 

50

100

51.10

98.57

102.20

98.57

 

50

100

49.95

101.24

99.90

101.24

X'

    

101.07

98.37

± S.D.

    

1.15

2.97

%RSD

    

1.13

3.02

%Error

    

0.65

1.74

2- Real plasma sample

100

100

91.87

92.48

91.87

92.48

 

100

100

90.12

93.76

90.12

93.76

 

100

100

86.25

87.98

86.25

87.98

X-

    

89.41

91.40

± S.D.

    

2.87

3.03

%RSD

    

3.21

3.321

%Error

    

1.85

1.91