Skip to main content

Table 6 Statistical parameters of ANOVA for the HPTLC method at 300 ng/band

From: Stability indicating high performance thin layer chromatography method development and validation for quantitative determination of tetracycline hydrochloride in tetracycline hydrochloride active pharmaceutical ingredient (API) and its dosage forms

S.No.

Parameters

Mobile phase composition

Different factors effect

Peak area

Rf

Peak area

Rf

1.

Model p-value

0.0747

0.1984

0.2019

0.1489

2.

Lack of fit

0.9594

0.8720

0.9627

0.891

3.

C.V. %

1.65

5.59

3.88

8.01

4.

R-squared

0.7506

0.8130

0.4210

0.7632

5.

Adj R-squared

0.5428

0.4857

0.2039

0.4789

6.

Pred R-squared

0.4968

0.4140

0.1069

0.4128

7.

Adeq precision

6.5380

6.0426

4.8219

6.1978

8.

Equation

Peak area = 

 + 0.0054

− 0.0000A

− 0.0000B

 + 0.0000C

− 0.0001AB

− 0.0001AC

Rf = 

0.2898

− 0.015 A

− 0.0105B

 + 0.0067C

 + 0.0077D

 + 0.005AB

− 0.0077AC

− 0.0067AD

Peak area = 

 + 0.0051

 + 0.0001B’

− 0.0001A’C’

 + 0.0001A’D’

Rf = 

 + 0.29

− 0.0075A’

− 0.042B’

− 0.005D’

− 0.0167A’B’

− 0.0075A’C’

− 0.015A’D’

  1. aA-ethyl acetate volume, B-acetone volume, C-methanol volume, D-1% Ammonia solution volume, A’-mobile phase volume, B’- tank saturation time, C’-activation time, D’- solvent migration distance